Leading Healthcare Solutions

FulCare
Water Treatment

Hemodialysis RO Water Treatment System

Hemodialysis RO water systems from 120 to 5,500 L/h, designed to meet YY0572-2015 for dialysis water, with 316L zero dead-leg circulation, IoT monitoring and optional heat disinfection.

CategoryWater Treatment
AvailabilityQuotation required
SupportOEM / ODM available
KEY FEATURES

Dialysis water with low bacteria and endotoxin

  • ROII 600–5,500 L/h and portable DVP-120 L/h models
  • Bacteria ≤1 cfu/mL and endotoxin ≤0.015 EU/mL for ROII
  • DVP-120: bacteria ≤50 cfu/mL and endotoxin ≤0.25 EU/mL
  • 316L stainless-steel zero dead-leg circulation loop
  • Optional heat disinfection and chemical disinfection
  • Water hardness and residual-chlorine monitoring options
  • PLC with full-colour touchscreen and remote IoT monitoring
  • PP anti-corrosion housing and duplex softener options
  • Direct feed to dialysis machines with stable pressure and flow

FulCare hemodialysis RO systems deliver water for 1–120 dialysis machines. The ROII series uses a 316L zero dead-leg circulation loop, PLC control and optional heat disinfection; the portable DVP-120 provides a compact solution for isolated or home dialysis.

120–5,500 L/hROII / DVP range
≤0.015 EU/mLEndotoxin target (ROII)
316LZero dead-leg loop
YY0572-2015Dialysis water standard

Final configuration depends on dialysis machine count, water analysis and room layout. Confirm before installation.

TECHNICAL SPECIFICATIONS

Configuration at a glance

SeriesROII-600 / 1000 / 1500 / 2500 / 3000 / 4500 / 5500; DVP-120
Capacity600–5,500 L/h (ROII); 120 L/h (DVP-120)
Dialysis machines5–10 / 10–18 / 18–30 / 30–50 / 50–70 / 90–100 / 90–120; 1–2 (DVP-120)
Water qualityROII: bacteria ≤1 cfu/mL, endotoxin ≤0.015 EU/mL; DVP-120: ≤50 cfu/mL, ≤0.25 EU/mL
StandardYY0572-2015 dialysis water
Loop316L stainless steel, zero dead-leg
ControlPLC + colour touchscreen; IoT remote monitoring
DisinfectionHeat / chemical, optional
MonitoringConductivity, pressure, flow, temperature; hardness / chlorine optional
HousingPP anti-corrosion
APPLICATIONS
Hemodialysis centres; hospital dialysis units; isolated dialysis rooms; home hemodialysis; renal care